Q 1. What
are the basic components for a HPLC system?
ANS - workable HPLC system must include a sampler, a
pump, a column, a detector, and date processor (computer
with software). An degasser and a column oven may also
be used for better quality of analysis. Basically there
are two types of of desigas. One is integrated design
which build all the components into one box. The other
type is modular design which allows swap and change of
components after installation.
Q 2. There are so many brands of HPLC available in the
market, how do I judge which one is the best for me?
ANS - Although the specification and price are
important, they should not be the only standard For many
chemists, the meaning of the specification may not be
easy to understand. Sometimes the sales person may
mislead you using their specification. Here are some
other items you may consider: a) feedback from other
users, b) hands-on experience or an instrument demo, c)
support capability of the vendor, d) cost of
maintenance, e) possibility of future upgrade.
Q 3. What are the important specifications for a
sampler?
ANS - Reproducibility, range of injection volume,
linearity over the injection range, and carry over. A
good linearity is very helpful when your sample
concentration varies a lot or when you want to make a
calibration curve using only one or two standard
solutions. The carry over property is very critical in
trace analysis. Sometimes a very complicated washing
program has to be used before each sample ran to avoid
carry over problem.
Q 4. What are the difference between a high pressure
gradient pump and a low pressure gradient pump?
ANS - In a low pressure gradient pump, different mobile
phase is mixed using a valve before entering the pump
head, m a Mgh pressure gradient pump, each different
mobile phase is delivered by an individual pump head and
then mixed at the pump outlet The low-pressure gradient
'design is of lower cost as it only use one pump head.
It can also ase more types of mobile phase without
significant increase of cost. The high-pressure gradient
design provides a much faster gradient since solvent
mixing jpoint is much closer to the column head. This is
measured using delay volume. The value can be 50-300 uL
for high pressure gradient pump and can be 2 to 3 times
larger for a low pressure gradient pump. A small delay
volume is important when the analysis time is short or
the flow rate is low. If the delay volume is too large,
it become impossible to obtain reproducible gradient run
since the planed composition cannot reach the column
head before a run is finished..
Q 5. What are the important specifications for an
isocratic pump?
ANS - Maximum working pressure, flow rate range, flow
reproducibilify and accuracy, pressure pulsation,
pressure monitoring and over pressure shut-off function.
Q 6. What are the important specifications for a
gradient pump?
ANS - Maximum working pressure, flow rate range, flow
reproducibility and accuracy, gradient repro^iueibility
and accuracy, delay volume, pressure pulsation, pressure
monitoring and over pressure shut-off function.
Q 7. Why is an on-line degasser always recommended for a
gradient pump?
ANS -Permanent gas has different solubility in
diflferent solvents. When different solvents are mixed
on line, the dissolved gas can be released from solvent
due to changed solubility. To verify this, you may add
some water to methanol and observe the large amount of
bubbles released from methanol. The bubble formed can
cause higher baseline nose and shift of retention time.
If you are doing an isocratic LC, a degasser is not so
critical since most bubble has been removed when you
filter the mobile phase under vacuum.
Q 8. How many types of detectors are available for HPLC?
ANS - UV detector, fluorescence detector,
electrochemical detector, conducti\dty detector,
refractive index detector, evaporative light scattering
detector, chiial detector, radioactive detector, mass
spectrometry detector (MSD).
Q 9. How to select a suitable detector according to my
application?
ANS -If your compounds absorb UV light within the range
above 200 nm, a UV type detector could be the best
choice. It is the most robust and most widely used HPLC
detector. Ehiorescence detector can have 10 to 100 times
better sensitivity for some compounds of large aromatic
ring in the siriictare (such as poly aromatic
hydrocarbons). Electrochemical detector is useful for
compounds that can be easily electrochemically reduced
or oxidized, such as phenols, aromatic amines, and some
carbohydrates. It can produce better method sensitivity
for such compounds than a UV detector due to better
selectivity. However, as the electrodes have direct
contact with column effluent and involve reaction with
sample matrix, the. response may not be as stable as a
UV detector. Conductivity detector is mainly for
inorganic ions in an ion exchange based liquid
chromatograpMc system. Normally a suppressor is needed
to remove background ions from mobile phase for a
satisfactory detection sensitivity. Refractive index
detector can detector any compounds but the sensitivity
is around 100-1000 times lower than a UV detector.
Another limit of this detector is that it cannot do
gradient LC. This detector is mainly used for organic
compounds that do not have reasonable UV absorbance,
such as small organic acids, carbohydrates, and some
polymers. Evaporative light scattering detector is of
similar application range as refractive index detector
and it can allow gradient HPLC. The disadvantage of this
detector is consumption of large amount of nitrogen for
evaporating the mobile phase, MSD is getting more and
more popular Mthough the price is 10-20 times higher
than other HPLC detectors. It can act as a general
purpose detector using scan mode or as an highly
selective detector using selective ion monitoring. The
most important advantage of MSD is in identity
confirmation.
Q 10. How to choose the flow cell volume for a UV
detector?
ANS - If the flow cell is too large, two adjacent peaks
may get mixed in the cell. If it is too small, the noise
may be higher due to less light reaching the photo
diodes. The higher back pressure could also limit the
flow rate range. The peak size may be used for the
selection. A narrow peak need a smaller flow cell. The
rule of thumb is the flow cell volume should not be more
than 1/3 of the peak volume. For example, if the peak
width is 0.1 min at 1 mL flow rate, the flow cell volume
should not be more than 33 uL. If your flow rate is 0.01
mL/min, the flow cell volume should be around 0.3 uL.
Q 11. How to choose an equivalent CIS column for my HPLC?
ANS - The situation in HPLC column is quite different
from GC columns. CIS or ODS columns from different
supplier can give very different elution profile. This
is because the silica properties, the carbon coverage
and the surface area varies considerably. If you want to
be sure of the same elution pattern without changing the
mobile phase, the column from the same manufacturer
should be used. If you want to try a column of similar
properties, you can look at columns that have similar
surface area and carbon loading for the packing
material.
Q 12. How to choose a guard column?
ANS - Its volume should not be more than 10% of the
analytical column. The packing should be similar to the
one in the analytical column. If this is not available,
you may use a guard column which has weaker retention to
your anarytes than the analytical column.
13. What are important specifications for a column?
Separation efficiency, inertness, durability, pH range,
and batch to batch reproducibility.
Q 14. How to choose a chiral column?
ANS - It is difficult to predict if a chiral column will
give good separation to a pair of chiral isomers. So the
best way is to search literature to see if it has been
done before. Or you may send your sample to the column
supplier for a sample testing,
Q 15. Some columns are end capped. What is it good for?
AANS - When the silica is bonded with C18, or C8, there
are still some active sites on the silica left over. An
end capping using a more active short chain can make the
column more inert to some basic compounds and make the
column more resistant to basic dissolution to the suica
material. However such capping is not stable at low pH
(<3). So normally such column is not very suitable for
mobile phase of low pH.
Q 16. How to make a column last longer?
Filter your sample and mobile phase. Make sure the pH of
the mobile phase is within the working range of the
column. Flush the column with methanol or acetonitrile
if it is not going to be used for sometime. |